Syringe diameter — gravimetric

Steps 2–5: weigh the delivered water → corrected diameter. 1 mg of water ≈ 1 µL. Do the pump mechanics first.

Weighed values → corrected diameter

Corrected diameter → set on the pump
— mm
fill in the weighed values above
Real flow (from balance)—
Flow error vs set point—
Q_real = mass / (ρ·t)  ·  D_real = D₀·√(Q_real / Q_set)

Gravimetric = ISO 4787-style; needs only an analytical balance. Verify linearity over ≥3 flow points (×3 each). A flow sensor is an optional cross-check. Works offline — no install.