Pump mechanics — plunger travel

Step 1 (do this first): does the drive head move the distance it should? Need the syringe step instead? → Syringe diameter (gravimetric).

Measured travel → lead-screw correction

Expected travel (what the pump should move)
— mm
set Q, time and diameter above
Mechanical error vs expected—
L_set = Q·t / (π·(D/2)²)  ·  L_real = |end − start|  ·  THREADROD_corr = THREADROD·(L_real / L_set)

Run ≥ 15 min (the longer it runs, the smaller the % reading error). L_real = |end − start|.

Mechanics = EURAMET CG-27 §4.3.3 displacement method (dial indicator / caliper). Before measuring, set gearbox · rod · microstep to match this pump. This lab's pumps use a T8 lead screw, 2.0 mm pitch (firmware stores it in tenths, so THREADROD = 20); ungeared → gearbox 1:1, geared → 100:1.